General observations: This Thermo Scientific TSX40086A ultra-low freezer is a clean unit in beautiful cosmetic condition and is now fully operational. We acquired it from a premier state facility during a department closure. Prior to release, the unit was processed under standard decontamination protocols in accordance with state guidelines.
Facility Testing: We first reviewed the freezer logs and noted an unexpected warming event in which temperatures settled around -26°C for an extended period before the unit was taken offline. Upon inspection, our technician found that the front intake filter had been removed, allowing the heat exchanger to become heavily blanketed with lint. The filter was replaced, and the freezer was serviced and thoroughly de-linted before resuming testing. During a 48-hour post-service test run, the unit showed normal pull-down, stable -80°C hold, and clean recovery through routine compressor off-cycles across the full observed window. All logged behavior during the test period was routine for an empty freezer operating to manufacturer spec. Test logs available upon request.
Additional notes: The interior is clean and sanitized per state release standards. All interface controls operated properly during testing. WiFi connected to our systems easily and we were able to establish admin user status. Door and seals are in excellent condition and the unit runs quietly during operation. The alarm battery currently presents as healthy and indicates approximately 11 months of remaining life; however, because the unit had been offline for more than six months, we recommend replacing the battery before any clinical or other mission-critical use.
Included in listing: The freezer comes exactly as shown and includes the OEM heavy-duty power cable. No additional accessories, racks, inventory, monitoring peripherals, or other items are included unless clearly shown in the photos.
Product Details
The Thermo Scientific TSX40086A is a TSX Series upright ultra-low temperature freezer designed for laboratory cold storage applications. Thermo Fisher product literature identifies this model as a 400-box class upright unit with a temperature range of -50°C to -86°C and approximately 19.4 cu. ft. (549 L) of interior capacity. The TSX platform includes a touchscreen control interface with on-unit access to temperature data, events, and maps, along with USB export capability.
Features
- Fully serviced operation: Post-service 48-hour test data showed normal pull-down, stable -80°C hold, and routine recovery behavior.
- Touchscreen controls: View temperature data, events, and maps directly on the color touchscreen interface.
- Large-capacity ULT storage: Approximately 19.4 cu. ft. (549 L) interior volume with storage for up to 400 standard 2" boxes.
- USB data export: Temperature history and event data can be exported directly from the unit.
- Clean, well-presenting unit: Interior and exterior present very well, with the interior sanitized per state release standards.
Applications
- Long-term laboratory storage of temperature-sensitive materials within the manufacturer’s specified operating range.
- Suitable for research, laboratory, industrial, or institutional cold storage applications where a fully operational TSX platform is desired.
Specifications
- Manufacturer: Thermo Scientific
- Model: TSX40086A
- Type: Upright ultra-low temperature freezer
- Temperature range: -50°C to -86°C
- Capacity: 19.4 cu. ft. (549 L)
- 2" box capacity: 400
- Rack capacity: Up to 16 racks
- Electrical requirements: 115 V, 60 Hz
- Included accessory: OEM heavy-duty power cable
- Status: Serviced and tested; see description for battery recommendation prior to clinical use